Lactate — Class II medical device recall Z-1931-2016
Terminated recall by Beckman Coulter Inc., reported 2016-06-15, distributed nationwide. Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the follow
| Recall number | Z-1931-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter Inc., Brea, CA, United States |
| Recall initiated | 2016-04-13 |
| Classified | 2016-06-08 |
| Reported by FDA | 2016-06-15 |
| Terminated | 2016-06-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GH, GR, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KE, KR, KW, LB, LK, LT, MA, MX, MY, NG, NL, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TR, TW, UA, UY, VE, VN, ZA |
| Model numbers | OSR6X93 |
Product
Lactate, Catalog No. OSR6x93 Product Usage: Lactate OSR6x93 is intended as a System reagent for the quantitative determination of L-Lactate in human plasma and cerebrospinal fluid (CSF) on Beckman Coulter AU analyzers.
Reason for recall
Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1931-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.