PathFinder NXT Pivoting Percutaneous Rod Holder This device is an… — Class II medical device recall Z-1933-2012

Terminated recall by Zimmer, Inc., reported 2012-07-11, distributed nationwide. Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted r

Recall numberZ-1933-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer, Inc., Warsaw, IN, United States
Recall initiated2012-06-05
Classified2012-07-03
Reported by FDA2012-07-11
Terminated2013-02-14
DistributedNationwide (US)
Countries outside the USAU, DE, FR
Quantity77 units
Model numbers3564-1

Product

PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery. The intended use of the pivoting percutaneous rod holder is to attach to the rod to facilitate rod insertion, positioning, and orientation through the extender sleeves, under the fascia. The rod is secured to the holder by tightening a captive set screw against the rod end in the tip of the holder using the set screw driver. After positioning the rod and securing at least one pedicle screw closure top against the implanted rod, the holder is removed from the rod end by loosening the captive set screw with the set screw driver.

Reason for recall

Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw was broken and material missing. There were no reports of fragments of the set screw being left in the patient or that there was any patient or user injury, but these events we