S4 Cervical — Class II medical device recall Z-1934-2012
Terminated recall by Aesculap Implant Systems LLC, reported 2012-07-11, distributed nationwide. Aesculap Implant Systems, LLC has initiated a voluntary removal of the S4 Cervical (S4C) Occiput Revision Inst
| Recall number | Z-1934-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Aesculap Implant Systems LLC, Center Valley, PA, United States |
| Recall initiated | 2012-03-07 |
| Classified | 2012-07-03 |
| Reported by FDA | 2012-07-11 |
| Terminated | 2012-08-08 |
| Distributed | Nationwide (US) |
| Quantity | 15 |
Product
S4 Cervical (S4C) Revision Instruments (FW099R and FW101R). The S4C Revision Instruments assist in the removal of an occipital cervical screw during revision surgery.
Reason for recall
Aesculap Implant Systems, LLC has initiated a voluntary removal of the S4 Cervical (S4C) Occiput Revision Instrument that is part of the S4C Occipital Instrument Set. The instruments may not thread into the screw properly during a revision surgery due to tolerance issues between the two instruments.