5 — Class II medical device recall Z-1935-2013

Terminated recall by Synthes USA HQ, Inc., reported 2013-08-21, distributed nationwide. DePuy Synthes has issued a Medical Device Correction notice for the 5.0mm Stainless Steel rods. This product w

Recall numberZ-1935-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSynthes USA HQ, Inc., West Chester, PA, United States
Recall initiated2013-06-05
Classified2013-08-09
Reported by FDA2013-08-21
Terminated2015-09-21
DistributedNationwide (US)
Quantity492
Lot numbers1602596, 1602617, 1880489, 2002330, 3000595, 3016224, 3080451, 3093619, 4729951, 4729952, 4729953, 4729954, 4729956, 4729958, 4729959, 4729960, 4729961, 4729962, 4835367, 4835368, 4835378, 4841209, 4841210, 4841211, 4841212, 4841213, 4856246, 4856248, 4856249, 4856250, 4856252, 4856275, 4919162, 4919165, 4923615, 4923651, 4923652, 4923653, 4923654, 4923655 and 43 more
Model numbers298.269, 298.270, 298.271, 298.272, 298.273, 298.274, 298.275, 298.276, 298.277, 298.278

Product

5.0mm Unit Rod 270mm 5.0mm Unit Rod 290mm 5.0mm Unit Rod 310mm 5.0mm Unit Rod 330mm 5.0mm Unit Rod 350mm 5.0mm Unit Rod 370mm 5.0mm Unit Rod 390mm 5.0mm Unit Rod 410mm 5.0mm Unit Rod 430mm 5.0mm Unit Rod 450mm pedicle screw spinal system; Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.

Reason for recall

DePuy Synthes has issued a Medical Device Correction notice for the 5.0mm Stainless Steel rods. This product was produced using a finishing process not identified as part of the manufacturing specification. No adverse event reports associated with the use of this product were received, and an extensive engineering and medical safety review concluded that the product poses no incremental risk to t