Data Foundry

Medical device recalls 2021

ONETOUCH ULTRA BLUE TEST STRIPS 100 COUNT Item Number: 2990174 — Class II medical device recall Z-1936-2021

Ongoing recall by Cardinal Health Inc., reported 2021-06-30, distributed nationwide. Due to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 h

Recall numberZ-1936-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCardinal Health Inc., Dublin, OH, United States
Recall initiated2021-03-16
Classified2021-06-23
Reported by FDA2021-06-30
DistributedNationwide (US)
Reason classestemperature control
UPC / GTIN / UDI-DI00353885009713
Model numbers2990174

Product

ONETOUCH ULTRA BLUE TEST STRIPS 100 COUNT Item Number: 2990174

Reason for recall

Due to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.