Data Foundry

Medical device recalls 2020

Arterial Line Kit 20ga X 3" — Class II medical device recall Z-1937-2020

Terminated recall by Argon Medical Devices, Inc, reported 2020-05-20, distributed nationwide. Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual st

Recall numberZ-1937-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmArgon Medical Devices, Inc, Athens, TX, United States
Recall initiated2020-04-08
Classified2020-05-08
Reported by FDA2020-05-20
Terminated2022-12-14
DistributedNationwide (US)
Quantity20
UPC / GTIN / UDI-DI00886333215577
Lot numbers11301012
Model numbers498106

Product

Arterial Line Kit 20ga X 3", Model 498106, Rx Only, Sterile EO, UDI: 00886333215577 - Product Usage: Arterial Catheters are indicated for use when in instances where continuous invasive blood pressure readings and/or frequent blood gas measurements are required. Case Label: Arterial Line Kit/1 20ga x 3", Qty: 10/CS, CE 2797

Reason for recall

Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.