Data Foundry

Medical device recalls 2020

PAC Tray/5 Catheter Introducer Tray 8F — Class II medical device recall Z-1938-2020

Terminated recall by Argon Medical Devices, Inc, reported 2020-05-20, distributed nationwide. Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual st

Recall numberZ-1938-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmArgon Medical Devices, Inc, Athens, TX, United States
Recall initiated2020-04-08
Classified2020-05-08
Reported by FDA2020-05-20
Terminated2022-12-14
DistributedNationwide (US)
Quantity80
UPC / GTIN / UDI-DI00886333214440
Lot numbers11299245
Model numbers497424

Product

PAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices. Curved tip Introducer allows for contralateral access to the iliac artery and selected vasculature. Case Label: PAC Tray/5 Catheter Introducer Tray 8F, QTY 10/Bx, Rx Only, STERILE EO

Reason for recall

Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1938-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.