Data Foundry

Medical device recalls 2016

Maestro Air Motor Pneumatic cranial drill motor — Class II medical device recall Z-1939-2016

Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2016-06-15, distributed in AL, AR, CA, FL, GA, IL, IN, KS…. Due to the buildup of grease within the hose and motor, the product may require multiple attempts to start. Th

Recall numberZ-1939-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Instruments Div. of Stryker Corporation, Portage, MI, United States
Recall initiated2016-04-28
Classified2016-06-09
Reported by FDA2016-06-15
Terminated2017-03-24
DistributedAL, AR, CA, FL, GA, IL, IN, KS, KY, MD, MI, MO, NC, NM, OH, OK, SC, TN, TX, VA, WA, WI
Quantity189
Reason classespackaging
Model numbers1522406703, 1522406713, 1522406733, 1522406743, 1522406753, 1522406763, 1522406773, 1522406793, 1522406803, 1522406813, 1522406833, 1522406843, 1522406853, 1522406863, 1522406873, 1522406883, 1522406893, 1522406903, 1522406923, 1523018583, 1523018603, 1523018623, 1523018633, 1523018683, 1523018713 and 165 more

Product

Maestro Air Motor Pneumatic cranial drill motor. The Maestro Air Pneumatic System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories.

Reason for recall

Due to the buildup of grease within the hose and motor, the product may require multiple attempts to start. The grease may also migrate into the attachment end of the motor. There is a negligible possibility that the orange grease may leak from the attachment end of the motor and contaminate the surgical wound resulting in soft tissue inflammation.