illumipro-10 Incubator/Reader — Class II medical device recall Z-1941-2014

Terminated recall by Meridian Bioscience Inc, reported 2014-07-09, distributed in AL, AZ, CA, CO, FL, IA, IN, LA…. Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorr

Recall numberZ-1941-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMeridian Bioscience Inc, Cincinnati, OH, United States
Recall initiated2014-04-22
Classified2014-06-27
Reported by FDA2014-07-09
Terminated2014-06-30
DistributedAL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, WV
Countries outside the USAU
Quantity35 instruments
Serial numbersIP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683, IP2684, IP2685, IP2686, IP2687, IP2688, IP2689, IP2690, IP2691, IP2692, IP2693, IP2694, IP2695 and 11 more

Product

illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.

Reason for recall

Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1941-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.