The Synchro2 Guidewires are a steerable guidewire family with a… — Class II medical device recall Z-1941-2024

Ongoing recall by Stryker Neurovascular, reported 2024-06-12, distributed nationwide. Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guid

Recall numberZ-1941-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmStryker Neurovascular, Fremont, CA, United States
Recall initiated2024-04-18
Classified2024-05-31
Reported by FDA2024-06-12
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, BG, BH, BR, CA, CH, CN, CO, CZ, DE, DK, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, IE, IL, IN, IR, IS, JO, JP, KR, KW, LT, LU, MX, NL, NO, PH, PL, PT, RO, RU, SA, SE, SG, SI, SK, TH, TR, TW, ZA
Quantity373,246
UPC / GTIN / UDI-DI04546540688729, 04546540688736, 04546540688743, 04546540688750, 04546540688767, 04546540688774, 04546540688781, 04546540688798, 07613327459630, 07613327459647, 07613327459654, 07613327459661
Lot numbers0000063501, 0000063520, 0000063568, 0000063716, 0000063767, 0000063879, 0000063886, 0000063894, 0000063905, 0000063933, 0000063985, 0000064072, 0000064113, 0000064191, 0000064269, 0000064281, 0000064309, 0000064357, 0000064360, 0000064370, 0000064410, 0000064420, 0000064482, 0000064486, 0000064487, 0000064527, 0000064565, 0000064566, 0000064589, 0000064590, 0000064640, 0000064735, 0000064736, 0000064796, 0000064797, 0000064882, 0000064927, 0000064960, 0000064961, 0000065004 and 460 more

Product

The Synchro2 Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014 in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque.

Reason for recall

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1941-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.