Data Foundry

Medical device recalls 2024

Arial mobile application — Class II medical device recall Z-1943-2024

Ongoing recall by Securitas Healthcare LLC, reported 2024-06-12, distributed nationwide. The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to

Recall numberZ-1943-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSecuritas Healthcare LLC, Lincoln, NE, United States
Recall initiated2024-04-11
Classified2024-05-31
Reported by FDA2024-06-12
DistributedNationwide (US)
Countries outside the USAU, CA, ES, FR, GB, HK, QA
Quantity661 sites
UPC / GTIN / UDI-DI00840948100963
Model numbers54630, 54630-SMA, 54640

Product

Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Arial Emergency and Nurse Call System.

Reason for recall

The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events when the app restarts, experience difficulty in placing and maintaining staff-to-staff phone calls, and frequently appears offline while the app restarts.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1943-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.