Data Foundry

Medical device recalls 2012

MOSAIQ Oncology Information Systems interfaced to Elekta Treatment… — Class II medical device recall Z-1944-2012

Terminated recall by Impac Medical Systems Inc, reported 2012-07-18, distributed in CT, GA, NJ, NY, VA. Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes

Recall numberZ-1944-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmImpac Medical Systems Inc, Sunnyvale, CA, United States
Recall initiated2012-06-08
Classified2012-07-06
Reported by FDA2012-07-18
Terminated2013-06-19
DistributedCT, GA, NJ, NY, VA
Countries outside the USBE, DK, FR, GB
Quantity11 affected

Product

MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086

Reason for recall

Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1944-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.