Data Foundry

Medical device recalls 2013

CELLTRACKS AUTOPREP System — Class II medical device recall Z-1949-2013

Terminated recall by Veridex, LLC, reported 2013-08-21, distributed nationwide. Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another pati

Recall numberZ-1949-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmVeridex, LLC, Raritan, NJ, United States
Recall initiated2012-12-28
Classified2013-08-10
Reported by FDA2013-08-21
Terminated2015-08-11
DistributedNationwide (US)
Countries outside the USAU, CA, CN, DE, ES, FR, GB, IN, IT, JP, SG

Product

CELLTRACKS AUTOPREP System (Software Version 2.4.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks AutoPrep System software 2.4.1 release notes Pub. No. J40135EN issued 2012-03-26 page 2, the document states that a new feature unique to software 2.4.1 allows the used to modify a batch in a new way. The user has additional optional fields in the edit sample information dialog box that allows one to enter expanded sample and patient information. The user can also edit the patient management setting for selected samples if not all samples have the same designation.

Reason for recall

Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1949-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.