TM MONO TIB STR GRN SZ 5 — Class II medical device recall Z-1953-2016
Terminated recall by Zimmer Trabecular Metal Technology, Inc., reported 2016-06-22, distributed nationwide. Zimmer Biomet is initiating a recall of specific lots of foil pouched, sterile Trabecular Metal implants that
| Recall number | Z-1953-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer Trabecular Metal Technology, Inc., Parsippany, NJ, United States |
| Recall initiated | 2016-04-15 |
| Classified | 2016-06-10 |
| Reported by FDA | 2016-06-22 |
| Terminated | 2017-11-29 |
| Distributed | Nationwide (US) |
| Quantity | 19,884 total |
| Reason classes | sterility |
Product
TM MONO TIB STR GRN SZ 5 - 00588606510 TM MONO TIB STR GRN SZ 5 12MM - 00588606512 TM MONO TIB STR GRN SZ 5 14MM - 00588606514 TM MONO TIB STR GRN SZ 5 17MM - 00588606517 TM MONO TIB STR GRN SZ 6 10MM - 00588606610 TM MONO TIB STR GRN SZ 6 12MM - 00588606612 TM MONO TIB STR GRN SZ 6 14MM - 00588606614 TM MONO TIB STR GRN SZ 6 17MM - 00588606617 TM MONO TIB STR BLUE SZ 7 10MM - 00588606710 TM MONO TIB STR BLUE SZ 7 12MM - 00588606712 TM MONO TIB STR BLUE SZ 7 14MM - 00588606714 TM MONO TIB STR BLUE SZ 7 17MM - 00588606717 TM MONO TIB STR GRN SZ 5 10MM - 00588607510 TM MONO TIB STR GRN SZ 5 12MM - 00588607512 TM MONO TIB STR GRN SZ 5 14MM - 00588607514 TM MONO TIB STR GRN SZ 5 17MM - 00588607517
Reason for recall
Zimmer Biomet is initiating a recall of specific lots of foil pouched, sterile Trabecular Metal implants that were manufactured between April 2011 and October 2012. The foil pouches are part of a dual barrier system to provide sterile integrity.