Simplexa" Flu A/B & RSV Direct — Class II medical device recall Z-1958-2014

Terminated recall by Focus Diagnostics Inc, reported 2014-07-09, distributed nationwide. Focus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B

Recall numberZ-1958-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmFocus Diagnostics Inc, Cypress, CA, United States
Recall initiated2014-05-15
Classified2014-07-01
Reported by FDA2014-07-09
Terminated2014-12-31
DistributedNationwide (US)
Quantity614 kits
Reason classeslabeling, software
Lot numbers614
Model numbers510, MOL2650

Product

Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413  Simplexa" Flu A/B & RSV Direct. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections in humans and is not intended to detect influenza C.

Reason for recall

Focus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B & RSV Direct (MOL2650). Focus Diagnostics received some customer complaints of Simplexa Flu A/B & RSV Direct assays with sporadic false signals, which may result in a higher false results rate due to a potential software spectral matrix and Direct Amplification Disc storage.