Hamilton Medical Infant Flow Sensor — Class II medical device recall Z-1964-2015

Terminated recall by Hamilton Medical, Inc., reported 2015-07-08, distributed nationwide. An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode

Recall numberZ-1964-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHamilton Medical, Inc., Reno, NV, United States
Recall initiated2015-06-01
Classified2015-07-02
Reported by FDA2015-07-08
Terminated2015-08-12
DistributedNationwide (US)
Quantity154 total
Model numbers155500

Product

Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ventilators.

Reason for recall

An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.