epoc BGEM Test Card — Class II medical device recall Z-1971-2013
Terminated recall by Epocal, reported 2013-08-21, distributed nationwide. This recall is being initiated because the affected lots of this product may yield results that are not consis
| Recall number | Z-1971-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Epocal, Ottawa, Ontario, Canada |
| Recall initiated | 2013-03-13 |
| Classified | 2013-08-14 |
| Reported by FDA | 2013-08-21 |
| Terminated | 2016-08-31 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AU, CA, CN, CO, DE, EG, GT, IL, IN, JP, KR, MX, MY, NL, NP, PE, PH, RU, SG, TH, VN, ZA |
| Quantity | 828,750 test |
| Lot numbers | 07-12283-00, 07-12284-00, 07-12285-00, 07-12289-00, 07-12290-00, 07-12292-00, 07-12293-00, 07-12296-00, 07-12297-00, 07-12298-00, 07-12300-00, 07-12303-00, 07-12305-00, 07-12306-00, 07-12307-00, 07-12310-00, 07-12312-00, 07-12313-00, 07-12314-00, 07-12317-00, 07-12320-00, 07-12321-00, 07-12324-00, 07-12325-00, 07-12327-00, 07-12331-00, 07-12333-00, 07-12335-00, 07-12338-00, 07-12340-00, 07-12341-00, 07-12345-00, 07-12346-00, 07-12347-00, 07-12349-00, 07-12352-00, 07-12354-00, 07-12355-00, 07-12362-00, 07-12363-00 and 10 more |
| Model numbers | CT-1004-00-00 |
Product
epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.
Reason for recall
This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test sensor (>400 mg/dl). Specifically, after a period of storage of three (3) to four (4) months, the Glucose sensor may report low results at the high end of the measurement range.