Data Foundry

Medical device recalls 2013

epoc BGEM Test Card — Class II medical device recall Z-1971-2013

Terminated recall by Epocal, reported 2013-08-21, distributed nationwide. This recall is being initiated because the affected lots of this product may yield results that are not consis

Recall numberZ-1971-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmEpocal, Ottawa, Ontario, Canada
Recall initiated2013-03-13
Classified2013-08-14
Reported by FDA2013-08-21
Terminated2016-08-31
DistributedNationwide (US)
Countries outside the USAR, AU, CA, CN, CO, DE, EG, GT, IL, IN, JP, KR, MX, MY, NL, NP, PE, PH, RU, SG, TH, VN, ZA
Quantity828,750 test
Lot numbers07-12283-00, 07-12284-00, 07-12285-00, 07-12289-00, 07-12290-00, 07-12292-00, 07-12293-00, 07-12296-00, 07-12297-00, 07-12298-00, 07-12300-00, 07-12303-00, 07-12305-00, 07-12306-00, 07-12307-00, 07-12310-00, 07-12312-00, 07-12313-00, 07-12314-00, 07-12317-00, 07-12320-00, 07-12321-00, 07-12324-00, 07-12325-00, 07-12327-00, 07-12331-00, 07-12333-00, 07-12335-00, 07-12338-00, 07-12340-00, 07-12341-00, 07-12345-00, 07-12346-00, 07-12347-00, 07-12349-00, 07-12352-00, 07-12354-00, 07-12355-00, 07-12362-00, 07-12363-00 and 10 more
Model numbersCT-1004-00-00

Product

epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.

Reason for recall

This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test sensor (>400 mg/dl). Specifically, after a period of storage of three (3) to four (4) months, the Glucose sensor may report low results at the high end of the measurement range.