Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley… — Class II medical device recall Z-1991-2020
Terminated recall by Covidien Llc, reported 2020-05-20, distributed nationwide. Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0
| Recall number | Z-1991-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Covidien Llc, Mansfield, MA, United States |
| Recall initiated | 2020-04-30 |
| Classified | 2020-05-14 |
| Reported by FDA | 2020-05-20 |
| Terminated | 2023-12-14 |
| Distributed | Nationwide (US) |
| Quantity | 46,828 units |
| Lot numbers | 1828205964, 1911414164 |
Product
Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, Premium, Urine Meter, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code P4P16XTSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Reason for recall
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment