Data Foundry

Medical device recalls 2025

ICU Medical Plum Solos — Class II medical device recall Z-1991-2025

Ongoing recall by ICU Medical, Inc., reported 2025-06-25, distributed nationwide. Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypa

Recall numberZ-1991-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmICU Medical, Inc., Lake Forest, IL, United States
Recall initiated2025-04-22
Classified2025-06-18
Reported by FDA2025-06-25
DistributedNationwide (US)
Countries outside the USPH
Quantity214 units
Reason classessoftware
Lot numbers60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051 and 174 more
Model numbers400011001

Product

ICU Medical Plum Solos, Item number 400011001; infusion pump

Reason for recall

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1991-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.