Infusomat — Class I medical device recall Z-1992-2012
Terminated recall by B. Braun Medical, Inc., reported 2012-07-25, distributed nationwide. Firm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the
| Recall number | Z-1992-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | B. Braun Medical, Inc., Allentown, PA, United States |
| Recall initiated | 2012-03-23 |
| Classified | 2012-07-16 |
| Reported by FDA | 2012-07-25 |
| Terminated | 2014-11-03 |
| Distributed | Nationwide (US) |
| Quantity | 38,225 pumps |
| Reason classes | device malfunction |
| Model numbers | 8713050U, 8713050U-99, 8713050US, 8713060U, 8713060U-99 |
Product
Infusomat(R) Space Pump (US Version) with software version 686G030103 (commonly termed 'G03'), 686E030003, 686F030005, 686F030007, 686G030002, and 686G030102. Intended for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.
Reason for recall
Firm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the pump door, when the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed. If the clip catch is broken and the door opened, free flow protection is still ensured. However, after the set is manually removed from the pump without the closure of t