Data Foundry

Medical device recalls 2013

Getinge Assure Accufast Biological Indicators — Class II medical device recall Z-1994-2013

Terminated recall by Getinge USA Inc, reported 2013-08-28. AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This cou

Recall numberZ-1994-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGetinge USA Inc, Rochester, NY, United States
Recall initiated2013-08-02
Classified2013-08-21
Reported by FDA2013-08-28
Terminated2017-02-23
Countries outside the USCA
Lot numbersR-430
Model numbers61301606636
Expiration dates2014-02-06

Product

Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623. Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.

Reason for recall

AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization parameters.