00595202010 Articular Surface XLPE CR ART SURF 1 — Class II medical device recall Z-2016-2013
Terminated recall by Zimmer, Inc., reported 2013-08-28, distributed nationwide. Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations
| Recall number | Z-2016-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer, Inc., Warsaw, IN, United States |
| Recall initiated | 2013-06-07 |
| Classified | 2013-08-22 |
| Reported by FDA | 2013-08-28 |
| Terminated | 2014-12-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | JP |
| Quantity | 192,355 all |
| Reason classes | labeling, unapproved product |
Product
00595202010 Articular Surface XLPE CR ART SURF 1,2/PUR 10, Rx, Sterile; 00595202012 Articular Surface XLPE CR ART SURF 1,2/PUR 12, Rx, Sterile; 00595202014 Articular Surface XLPE CR ART SURF 1,2/PUR 14, Rx, Sterile; 00595202017 Articular Surface XLPE CR ART SURF 1,2/PUR 17, Rx, Sterile; 00595202020 Articular Surface XLPE CR ART SURF 1,2/PUR 20, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Reason for recall
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d