Data Foundry

Medical device recalls 2019

AXIOM Artis BA — Class II medical device recall Z-2016-2019

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2019-07-31, distributed nationwide. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100

Recall numberZ-2016-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2019-06-12
Classified2019-07-22
Reported by FDA2019-07-31
Terminated2020-09-18
DistributedNationwide (US)
Quantity7
Serial numbers28139, 28144, 28154, 28155, 28156, 28159, 28160
Model numbers5904656

Product

AXIOM Artis BA, Model Number 5904656

Reason for recall

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2016-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.