VITALIO DR EL — Class I medical device recall Z-2017-2021
Ongoing recall by Boston Scientific Corporation, reported 2021-07-21. The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initia
| Recall number | Z-2017-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Boston Scientific Corporation, Saint Paul, MN, United States |
| Recall initiated | 2021-06-03 |
| Classified | 2021-07-12 |
| Reported by FDA | 2021-07-21 |
| Quantity | 16,525 units |
| Serial numbers | 029, 053, 100, 100000, 100001, 100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100012, 100013, 100014, 100015, 100016, 100017, 100018, 100019, 100020, 100021 and 475 more |
Product
VITALIO DR EL (pacemaker), Model Numbers: a) J274 b) J277 c) K274 (added 7/15/2021) d) K277 (added 7/15/2021) e) K284 (added 7/15/2021) f) K287 (added 7/15/2021)
Reason for recall
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2017-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.