AXIOM Artis TA — Class II medical device recall Z-2018-2019

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2019-07-31, distributed nationwide. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100

Recall numberZ-2018-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2019-06-12
Classified2019-07-22
Reported by FDA2019-07-31
Terminated2020-09-18
DistributedNationwide (US)
Quantity5
Serial numbers32085, 32090, 32091, 32094, 32095
Model numbers7007755

Product

AXIOM Artis TA, Model Number 7007755

Reason for recall

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.