Data Foundry

Medical device recalls 2024

VITROS Immunodiagnostic Products Vitamin B12 Range Verifier — Class III medical device recall Z-2019-2024

Ongoing recall by Ortho-Clinical Diagnostics, Inc., reported 2024-06-12, distributed nationwide. The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2

Recall numberZ-2019-2024
Product typeMedical device
ClassificationClass III
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Inc., Rochester, NY, United States
Recall initiated2024-05-08
Classified2024-06-04
Reported by FDA2024-06-12
DistributedNationwide (US)
Quantity76 units
Reason classespotency
UPC / GTIN / UDI-DI10758750000937
Lot numbers1200
Expiration dates2024-09-30

Product

VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration range of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems when used for the quantitative measurement of vitamin B12 - Product Code: 6800872

Reason for recall

The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.