ARTIS is a family of dedicated angiography systems developed for… — Class II medical device recall Z-2019-2025
Ongoing recall by Siemens Medical Solutions USA, Inc, reported 2025-07-09. In very rare situations, communication between the sensor measuring dose applied to the area and the correspon
| Recall number | Z-2019-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | FDA mandated |
| Recalling firm | Siemens Medical Solutions USA, Inc, Malvern, PA, United States |
| Recall initiated | 2025-05-15 |
| Classified | 2025-06-27 |
| Reported by FDA | 2025-07-09 |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 04056869046877 |
Product
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e., patient extremities. This does not include projection radiography. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.
Reason for recall
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.