Aquatec Fixed Backrest — Class II medical device recall Z-2023-2014
Terminated recall by Invacare Corporation, reported 2014-07-16. The Aquatec¿ Fixed Backrest withstands a load of only 250 newtons at uneven load. This may result in the Fixed
| Recall number | Z-2023-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Invacare Corporation, Elyria, OH, United States |
| Recall initiated | 2014-06-25 |
| Classified | 2014-07-09 |
| Reported by FDA | 2014-07-16 |
| Terminated | 2015-04-15 |
| Countries outside the US | CA |
| Quantity | 465 units |
| Reason classes | device malfunction |
| Model numbers | 1470713, 1470793, 1525707, 1534327 |
Product
Aquatec Fixed Backrest - accessory for Mobile Shower and Toilet commode Aquatec Ocean, Ocean VIP and Ocean E-VIP family.
Reason for recall
The Aquatec¿ Fixed Backrest withstands a load of only 250 newtons at uneven load. This may result in the Fixed Backrest becoming detached from the frame.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2023-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.