Data Foundry

Medical device recalls 2012

OEC 9800 — Class II medical device recall Z-2026-2012

Terminated recall by GE OEC Medical Systems, Inc, reported 2012-07-25. Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues th

Recall numberZ-2026-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE OEC Medical Systems, Inc, Salt Lake City, UT, United States
Recall initiated2007-11-19
Classified2012-07-18
Reported by FDA2012-07-25
Terminated2013-01-28
Quantity11,100 units
Serial numbers9800
Model numbers00-884777-01, 00-884909-01, 00-884911-01, 00-884913-01, 00-885553-01, 00-885554-01, 00-885555-01, 00-886289-01

Product

OEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, OEC 9800 MD Motorized C-arm System designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and is also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may be used for other imaging applications at the physician's discretion.

Reason for recall

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2026-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.