Data Foundry

Medical device recalls 2014

The Hospira MedNet Medication Management Suite — Class II medical device recall Z-2027-2014

Terminated recall by Hospira Inc., reported 2014-07-16, distributed nationwide. Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "mi

Recall numberZ-2027-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHospira Inc., Lake Forest, IL, United States
Recall initiated2014-06-06
Classified2014-07-09
Reported by FDA2014-07-16
Terminated2016-12-06
DistributedNationwide (US)
Countries outside the USHK
Quantity13 units
Reason classessoftware

Product

The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters. The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physicians orders.

Reason for recall

Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2027-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.