Data Foundry

Medical device recalls 2013

90597004109 Articular Surface CR ART SURF AE56/STR GRN 09MM — Class II medical device recall Z-2031-2013

Terminated recall by Zimmer, Inc., reported 2013-08-28, distributed nationwide. Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations

Recall numberZ-2031-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer, Inc., Warsaw, IN, United States
Recall initiated2013-06-07
Classified2013-08-22
Reported by FDA2013-08-28
Terminated2014-12-15
DistributedNationwide (US)
Countries outside the USJP
Quantity192,355 all
Reason classeslabeling, unapproved product

Product

90597004109 Articular Surface CR ART SURF AE56/STR GRN 09MM, Rx, Sterile; 90597004110 Articular Surface CR ART SURF AE56/STR GRN 10MM, Rx, Sterile; 90597004112 Articular Surface CR ART SURF AE56/STR GRN 12MM, Rx, Sterile; 90597004114 Articular Surface CR ART SURF AE56/STR GRN 14MM, Rx, Sterile; 90597004117 Articular Surface CR ART SURF AE56/STR GRN 17MM, Rx, Sterile; 90597004120 Articular Surface CR ART SURF AE56/STR GRN 20MM, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Reason for recall

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2031-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.