Data Foundry

Medical device recalls 2013

00597204109 Articular Surface CR ART SURF 56/STRIPE GRN 9 — Class II medical device recall Z-2032-2013

Terminated recall by Zimmer, Inc., reported 2013-08-28, distributed nationwide. Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations

Recall numberZ-2032-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer, Inc., Warsaw, IN, United States
Recall initiated2013-06-07
Classified2013-08-22
Reported by FDA2013-08-28
Terminated2014-12-15
DistributedNationwide (US)
Countries outside the USJP
Quantity192,355 all
Reason classeslabeling, unapproved product

Product

00597204109 Articular Surface CR ART SURF 56/STRIPE GRN 9, Rx, Sterile; 00597204110 Articular Surface CR ART SURF 56/STRIPE GRN 10, Rx, Sterile; 00597204112 Articular Surface CR ART SURF 56/STRIPE GRN 12, Rx, Sterile; 00597204114 Articular Surface CR ART SURF 56/STRIPE GRN 14, Rx, Sterile; 00597204117 Articular Surface CR ART SURF 56/STRIPE GRN 17, Rx, Sterile; 00597204120 Articular Surface CR ART SURF 56/STRIPE GRN 20, Rx, Sterile; 00597204123 Articular Surface CR ART SURF 56/STRIPE GRN 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Reason for recall

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d