Data Foundry

Medical device recalls 2025

Hillrom LikoScale component — Class I medical device recall Z-2032-2025

Ongoing recall by Baxter Healthcare Corporation, reported 2025-07-16. The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling

Recall numberZ-2032-2025
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corporation, Deerfield, IL, United States
Recall initiated2025-05-30
Classified2025-07-09
Reported by FDA2025-07-16
Quantity9,839 units
Reason classesdevice malfunction

Product

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Reason for recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2032-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.