HeartStart MRx Monitor/Defibrillator — Class II medical device recall Z-2035-2018

Terminated recall by Philips Electronics North America Corporation, reported 2018-06-06, distributed in NC, OH. The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been lo

Recall numberZ-2035-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Electronics North America Corporation, Andover, MA, United States
Recall initiated2018-05-03
Classified2018-05-30
Reported by FDA2018-06-06
Terminated2019-10-17
DistributedNC, OH
Countries outside the USLB
Quantity5
Model numbersM3535A

Product

HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillators (non-wearable) Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

Reason for recall

The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.