Alere Triage Tox+MTD Drug Screen — Class II medical device recall Z-2039-2012
Terminated recall by Alere San Diego, Inc., reported 2012-07-25, distributed nationwide. A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have signific
| Recall number | Z-2039-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Alere San Diego, Inc., San Diego, CA, United States |
| Recall initiated | 2012-05-14 |
| Classified | 2012-07-19 |
| Reported by FDA | 2012-07-25 |
| Terminated | 2012-12-04 |
| Distributed | Nationwide (US) |
| Quantity | 9,997 kits |
Product
Alere Triage Tox+MTD Drug Screen, PN 94400 Product Usage: The Alere Triage TOX Drug Screen is a fluorescence immunoassay to be used with the Alere Triage Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses. This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/ Mass Spectrometry/ Mass Spectrometry (LC/MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods. A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.
Reason for recall
A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has been observed for Triage TOX Drug Screen.