Brand Name: VITROS Product Name: VITROS 4600 Chemistry System… — Class II medical device recall Z-2040-2025
Ongoing recall by Ortho-Clinical Diagnostics, Inc., reported 2025-07-09, distributed nationwide. a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the s
| Recall number | Z-2040-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Inc., Rochester, NY, United States |
| Recall initiated | 2025-05-09 |
| Classified | 2025-06-27 |
| Reported by FDA | 2025-07-09 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 10758750012343, 10758750033201 |
| Serial numbers | 46000111-46001897 |
| Model numbers | 6802445, 6900440 |
Product
Brand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog Number: (1) 6802445, (2) 6900440 Software Version: version 3.2 to version 3.8.3 (all available software versions) Product Description: In vitro diagnostic chemistry analyzer.
Reason for recall
a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.