Data Foundry

Medical device recalls 2024

MRI systems: Vantage Orian MRT-1550 — Class II medical device recall Z-2041-2024

Ongoing recall by Canon Medical System, USA, INC., reported 2024-06-19, distributed in AL, AR, AZ, CA, CO, CT, FL, GA…. The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secur

Recall numberZ-2041-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCanon Medical System, USA, INC., Tustin, CA, United States
Recall initiated2024-05-10
Classified2024-06-09
Reported by FDA2024-06-19
DistributedAL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MT, NH, NJ, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, VT, WI, WV, WY
Quantity153
UPC / GTIN / UDI-DI04987670101632, 04987670101649, 04987670102400, 04987670102615, 04987670103360
Serial numbersMRT-1550

Product

MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)

Reason for recall

The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2041-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.