Crit Line in a Clip — Class II medical device recall Z-2044-2016

Terminated recall by Fresenius Medical Care Renal Therapies Group, LLC, reported 2016-06-29, distributed nationwide. Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result i

Recall numberZ-2044-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmFresenius Medical Care Renal Therapies Group, LLC, Waltham, MA, United States
Recall initiated2016-05-24
Classified2016-06-21
Reported by FDA2016-06-29
Terminated2021-04-30
DistributedNationwide (US)
Quantity875 units
Model numbers190895, 2008TGEN

Product

Crit Line in a Clip (CLiC). CliC P/N: CL 10041001 CLiC is a continuous real-time monitor for non-invasive hematocrit oxygen saturation and percent change in blood volume measurement during hemodialysis treatment. Used in conjunction with the Fresenius 2008T Hemodialysis Machines as follows: Model: 190895 2008T GEN 2 BIBAG w/o CDX 190713 2008T Hemodialysis SYS with CDX 190858 2008T Hemodialysis System w/o CDX 190766 2008T Hemodialysis System w/o CDX

Reason for recall

Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop.