Germ Terminator — Class II medical device recall Z-2047-2016
Terminated recall by Ceg Enterprises Llc, reported 2016-06-29, distributed nationwide. CEG Enterprises, LLC is recalling Germ Terminator Toothbrush Sanitizer GT100 due to the following issues: UL l
| Recall number | Z-2047-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ceg Enterprises Llc, Hopkins, MN, United States |
| Recall initiated | 2016-04-05 |
| Classified | 2016-06-23 |
| Reported by FDA | 2016-06-29 |
| Terminated | 2016-11-22 |
| Distributed | Nationwide (US) |
| Quantity | 100,548 units |
| Reason classes | labeling, specification failure |
| Model numbers | GT100 |
Product
Germ Terminator, Toothbrush Steam Sterilization for Oral Health. Model GT100. The Germ Terminator uses steam heat to sanitize two manual toothbrushes or two toothbrush heads. Water is poured into a reservoir in the device's housing. The steam cycle is followed by a drying cycle.
Reason for recall
CEG Enterprises, LLC is recalling Germ Terminator Toothbrush Sanitizer GT100 due to the following issues: UL label believed to be not genuine; discoloration of power cords (could indicate some degradation); reports of excessive heat.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2047-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.