EcoFit cup cementless: Model — Class II medical device recall Z-2050-2021
Terminated recall by IMPLANTCAST GMBH, reported 2021-07-21, distributed nationwide. A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, whic
| Recall number | Z-2050-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | IMPLANTCAST GMBH, Buxtehude, Germany |
| Recall initiated | 2021-05-10 |
| Classified | 2021-07-09 |
| Reported by FDA | 2021-07-21 |
| Terminated | 2022-11-18 |
| Distributed | Nationwide (US) |
| Quantity | 41 units |
| UPC / GTIN / UDI-DI | 04048844005368, 04048844005375, 04048844005382, 04048844005399, 04048844005405, 04048844005412, 04048844005429, 04048844005436, 04048844005443, 04048844005450, 04048844005467, 04048844005474 |
| Lot numbers | 02200048, 02200050, 02200052, 02200054, 02200056, 02200058, 02200060, 02200062, 02200064, 02200066, 02200068, 1416I6G089, 1439O4K008, 15190KC048, 1736252001, 1736252003, 1736252004, 1736252006, 1736252007, 1736252008, 1736252010, 1736253002, 1736253004, 1736253005, 1736253006, 1736253007, 1736253008, 1736253014, 1736253015, 1736253016, 1736253017, 1736253019, 1736253020, 1736253021, 1736253022, 1736253023, 1738104009, 173810D001, 173810D002, 173810D003 and 197 more |
| Model numbers | 02200046 |
Product
EcoFit cup cementless: Model(Catalog Number)/Description: 02200046 /EcoFit cup cementless 46mm dia incl. central hole cover; 02200048 /EcoFit cup cementless 48mm dia incl. central hole cover; 02200050 /EcoFit cup cementless 50mm dia incl. central hole cover; 02200052 /EcoFit cup cementless 52mm dia incl. central hole cover; 02200054 /EcoFit cup cementless 54mm dia incl. central hole cover; 02200056 /EcoFit cup cementless 56mm dia incl. central hole cover; 02200058 /EcoFit cup cementless 58mm dia incl. central hole cover; 02200060 /EcoFit cup cementless 60mm dia incl. central hole cover; 02200062 /EcoFit cup cementless 62mm dia incl. central hole cover; 02200064 /EcoFit cup cementless 64mm dia incl. central hole cover; 02200066 /EcoFit cup cementless 66mm dia incl. central hole cover; 02200068 /EcoFit cup cementless 68mm dia incl. central hole cover;
Reason for recall
A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor.