Philips HeartStart XL Defibrillator/Monitor — Class II medical device recall Z-2053-2020

Terminated recall by Philips North America, LLC, reported 2020-05-27, distributed nationwide. The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device

Recall numberZ-2053-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, LLC, Andover, MA, United States
Recall initiated2020-04-13
Classified2020-05-20
Reported by FDA2020-05-27
Terminated2021-04-07
DistributedNationwide (US)
Quantity94,034 units
Reason classesdevice malfunction
Model numbersM4735A

Product

Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.

Reason for recall

The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.