BAM — Class II medical device recall Z-2055-2014
Terminated recall by Stryker Communications, reported 2014-07-23, distributed nationwide. During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if t
| Recall number | Z-2055-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Stryker Communications, Flower Mound, TX, United States |
| Recall initiated | 2014-03-17 |
| Classified | 2014-07-15 |
| Reported by FDA | 2014-07-23 |
| Terminated | 2015-05-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BR, CA, CL, CN, CO, JP, KR, MX, PH |
| Quantity | 953 shape |
| Serial numbers | 080924DC001 |
| Model numbers | 0682000976, P17087 |
Product
BAM (SHAPE) ARM, LEFT, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as component of kit part number 0682000976, and the Arm part number in the kit is P17087. The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.
Reason for recall
During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.