Data Foundry

Medical device recalls 2012

System 1 Base 220/240V The Terumo¿ Advanced Perfusion System 1 is… — Class II medical device recall Z-2056-2012

Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2012-08-01, distributed nationwide. Terumo Cardiovascular Systems (Terumo CVS) received eight complaints since September 2008 of drive motor failu

Recall numberZ-2056-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTerumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States
Recall initiated2012-06-22
Classified2012-07-20
Reported by FDA2012-08-01
Terminated2013-03-06
DistributedNationwide (US)
Countries outside the USAE, AR, BE, CA, CL, CO, CR, DE, DO, EG, HK, HN, ID, IL, IN, JO, JP, KR, KW, MX, MY, PH, PK, RU, SA, SG, TR, TW, VN, ZA
Quantity640 devices
Serial numbers0006-0066, 0100-0311, 0313-0322, 1001-1027, 1100-1429
Model numbers801764

Product

System 1 Base 220/240V The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Reason for recall

Terumo Cardiovascular Systems (Terumo CVS) received eight complaints since September 2008 of drive motor failure of the Sarns Centrifugal System of the Terumo Advanced Perfusion System 1. A failure of the drive motor for the Terumo System 1 centrifugal system triggers error messages and alarms; therefore it is easily recognizable by the user. The potential hazards are dependent on the application

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2056-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.