Data Foundry

Medical device recalls 2013

Synthes 2 — Class II medical device recall Z-2057-2013

Terminated recall by Synthes (USA) Products LLC, reported 2013-09-04, distributed nationwide. The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal

Recall numberZ-2057-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSynthes (USA) Products LLC, West Chester, PA, United States
Recall initiated2013-05-31
Classified2013-08-26
Reported by FDA2013-09-04
Terminated2015-09-21
DistributedNationwide (US)
Reason classeslabeling
Model numbersJ8301-F

Product

Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.

Reason for recall

The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm