Synthes 2 — Class II medical device recall Z-2057-2013
Terminated recall by Synthes (USA) Products LLC, reported 2013-09-04, distributed nationwide. The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal
| Recall number | Z-2057-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Synthes (USA) Products LLC, West Chester, PA, United States |
| Recall initiated | 2013-05-31 |
| Classified | 2013-08-26 |
| Reported by FDA | 2013-09-04 |
| Terminated | 2015-09-21 |
| Distributed | Nationwide (US) |
| Reason classes | labeling |
| Model numbers | J8301-F |
Product
Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
Reason for recall
The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm