Data Foundry

Medical device recalls 2015

Hamilton-G5 Ventilators with software versions V2 — Class II medical device recall Z-2057-2015

Terminated recall by Hamilton Medical, Inc., reported 2015-08-05, distributed nationwide. Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longe

Recall numberZ-2057-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHamilton Medical, Inc., Reno, NV, United States
Recall initiated2015-03-24
Classified2015-07-24
Reported by FDA2015-08-05
Terminated2015-12-23
DistributedNationwide (US)
Quantity248
Model numbers159001, 159002

Product

Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.

Reason for recall

Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2057-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.