Hamilton-G5 Ventilators with software versions V2 — Class II medical device recall Z-2057-2015
Terminated recall by Hamilton Medical, Inc., reported 2015-08-05, distributed nationwide. Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longe
| Recall number | Z-2057-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hamilton Medical, Inc., Reno, NV, United States |
| Recall initiated | 2015-03-24 |
| Classified | 2015-07-24 |
| Reported by FDA | 2015-08-05 |
| Terminated | 2015-12-23 |
| Distributed | Nationwide (US) |
| Quantity | 248 |
| Model numbers | 159001, 159002 |
Product
Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.
Reason for recall
Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2057-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.