Centurion Convenience Kits — Class I medical device recall Z-2067-2019
Terminated recall by Medline Industries Inc, reported 2019-08-21, distributed in IL. This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue wit
| Recall number | Z-2067-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medline Industries Inc, Northfield, IL, United States |
| Recall initiated | 2019-05-10 |
| Classified | 2019-08-12 |
| Reported by FDA | 2019-08-21 |
| Terminated | 2021-08-18 |
| Distributed | IL |
| Quantity | 4,890 |
| Reason classes | packaging, device malfunction |
| Lot numbers | 2018090590, 2018090690, 2018101190, 2018101590, 2018112190, 2018121190, 2019013190, 2019030490, 2019040190 |
Product
Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.
Reason for recall
This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2067-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.