IMPAX CV Reporting The Results Management — Class II medical device recall Z-2069-2012
Terminated recall by AGFA Corp., reported 2012-08-01, distributed nationwide. When users selected "Left stenosis" in the "Graft Duplex Conclusion" section in IMPAX CV Reporting Non-Invasiv
| Recall number | Z-2069-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | AGFA Corp., Greenville, SC, United States |
| Recall initiated | 2012-07-10 |
| Classified | 2012-07-24 |
| Reported by FDA | 2012-08-01 |
| Terminated | 2013-10-29 |
| Distributed | Nationwide (US) |
| Quantity | 6 |
Product
IMPAX CV Reporting The Results Management (RM)/IMPAX CV reporting component facilitates the quick, effective creating of digital structured reports for : adult catheterization and echocardiography, pediatric echocardiography, congenital heat disease, nuclear cardiology and non-invasive vascular disease management. This webb-based application allows cardiology departments to view, edit and sign the structured reports.
Reason for recall
When users selected "Left stenosis" in the "Graft Duplex Conclusion" section in IMPAX CV Reporting Non-Invasive(NIV) module, "Right stenosis" appeared in the sentence display within the printed representation of the report.