Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy… — Class II medical device recall Z-2070-2014
Terminated recall by Excelsior Medical Corp, reported 2014-07-23, distributed nationwide. There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow
| Recall number | Z-2070-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Excelsior Medical Corp, Neptune, NJ, United States |
| Recall initiated | 2014-06-05 |
| Classified | 2014-07-16 |
| Reported by FDA | 2014-07-23 |
| Terminated | 2015-08-19 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA |
| Reason classes | packaging |
| Lot numbers | 03030813, 03050813, 03091213, 64031513, 64073113, 64081313, 64090513 |
Product
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
Reason for recall
There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2070-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.