Getinge brand --- large capacity — Class II medical device recall Z-2079-2012

Terminated recall by Getinge USA Inc, reported 2012-08-01, distributed in IN, KS, MD, NM, TX. Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series

Recall numberZ-2079-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGetinge USA Inc, Rochester, NY, United States
Recall initiated2012-05-25
Classified2012-07-26
Reported by FDA2012-08-01
Terminated2016-05-18
DistributedIN, KS, MD, NM, TX
Countries outside the USCA, MX
Quantity8 washers
Serial numbersW50017381, W50020296, W50020952, W50024258, W50024259, W50025790, W50026573, W50028138

Product

Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128

Reason for recall

Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error.