Data Foundry

Medical device recalls 2025

REF: 666289 — Class II medical device recall Z-2079-2025

Ongoing recall by Becton, Dickinson and Company, BD Bio Sciences, reported 2025-07-09, distributed in AR, AZ, CA, CO, CT, DC, DE, FL…. Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accura

Recall numberZ-2079-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBecton, Dickinson and Company, BD Bio Sciences, Milpitas, CA, United States
Recall initiated2025-06-03
Classified2025-07-03
Reported by FDA2025-07-09
DistributedAR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WI, WV
Countries outside the USAR, AU, BE, BR, CA, CL, CN, CO, DO, EC, HK, IN, JP, KR, MX, MY, NP, NZ, PE, PK, SG, SV, TW
Quantity278 units
UPC / GTIN / UDI-DI00382906662897
Lot numbers85637, 85638
Model numbers666289

Product

REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.

Reason for recall

Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2079-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.